      Fluxes are hygroscopic, in particular lithium tetraborate. Therefore, the melting agents should be dried before use, and preheating at 105 °C over a longer period (overnight) is recommended.

       Other additives are sometimes also used in the melting process. These may be oxidizing agents and glass-forming agents, but also substances that aim to improve the quality of the fusion bead or have an influence on the analytical process.By strong absorbers, such as BaO or La2O3, the matrix interaction is increased and fusion beads can then be treated as infinitely thick samples, in which the intensity of the fluorescence signal does not depend on the sample thickness, i.e. their information depth is smaller.

       Sometimes, also lithium bromide (LiBr), lithium iodide (LiI), or ammonium iodide (NH4I) is added to the molten sample in small amounts (one or two drops) to increase the surface tension of the melt (especially when using lithium metaborate as flux). This allows for a loss-free pouring of the melt into the cooling mold and a subsequent simple separation of the sample from the mold. For these additives, care must be taken again that no interferences with analyte elements occur. Examples are overlaps of the Br-Lα line with the Al-Kα line or the Mg-Kα line, and the I-Lβ lines with the Ti-Kα line. However, the contents of Br and I in the fused bead are very small, also caused by their high volatility. At the melting point of lithium metaborate half of the initial amount of Br or I is evaporated within approximately eight or two minutes, respectively.

       A good mixture of the sample material, additives, and the melting agent is very important. It should be carried out in ball mills in order to prevent segregation of the different components. Mixing in the melting vessel does not lead to a sufficient result and can also disturb the surface of the crucible, which can prevent the complete pouring of the sample or lead to its early destruction.

       The oxidation and melting process usually take place above a gas flame, but electrically heated muffle furnaces are also used. In order to increase the productivity of the sample preparation, usually several samples can be processed simultaneously.A mound in the bottom of the Pt mold improves the mixing effect during the pivoting of the crucible.

       In order to increase the quality of the fusion beads and to standardize them, the temperature profiles of the ovens are mostly controllable. The temperatures for the melt process should be about 150–200 °C above the melting point of the melt. However, in order to reduce the loss of volatile elements such as alkali metals or sulfur, the melting temperatures should not be too high. This can be achieved using Na2B4O7 (see Table 3.9).

      Equipment for fusion melting is offered by companies such as Claisse Inc., Fluxana GmbH, or XRF Scientific Ltd. (Table B.1). All of them are temperature controlled, mix the melt automatically by pivoting or rotating the crucible during the melting process, and pour the melt into a preheated casting mold, which is then cooled by a predetermined temperature profile. The casting mold then already has the required size of the X-ray spectrometer being used. The companies also offer prepared recipes for the preparation of a wide variety of sample materials.

(Left) Image of the Pt crucibles over a burner during the melting process and (right) a typical temperature profile for the complete melting process. Graph depicting a cooling curve representing the influence of the cooling rate on the quality of glass beads.

      The manufacturing of fusion beads requires not only a relatively long time, but also a high manual effort. Various efforts have been made to automate this process (see for example Flock and Ohls 1987).

      3.4.4.3 Loss of Ignition

      During the melting process, the weight of the sample can be changed due to the loss of volatile elements or compounds, especially by the evaporation of moisture or gas inclusions in the sample material. This loss – LOI – must be taken into account in the later quantification. The following relationship is used:

      (3.2)equation

      3.4.4.4 Quality Criteria for Fusion Beads

      Fusion beads of high quality are characterized by good transparency, a flat and smooth surface, and high homogeneity. Depending on the glass former, the fusion beads can be hygroscopic, i.e. their storage in desiccators is recommended if the samples should be available for follow-up measurements.

Examples of a homogeneous (a) as well as a milky and cracked (b) fusion bead that was cooled too quickly.

      The following defects can occur in fusion beads:

       Crystallization can occur when a non-suitable flux for the analyzed material is used or when the cooling rate of the beads is too slow.

       Cracking occurs if the internal stress in the bead is too high. These can be caused by unsuitable annealing procedures or due to small inclusions, for example, after crystallization in the glassy sample.

       Bubbles can be generated if the material contains compounds that are volatile, if the material was not sufficiently dried, if the melt is too viscous, or if the melt captures air inclusions during the pouring process.

       Distortion, in cases where the mold has deviations from the correct shape these deviations are then transferred to the fusion bead and can lead, for example, to uneven surfaces.

      All these influences must be tested and optimized for the various sample materials as well as for the available equipment.