Эротические рассказы

Essential Concepts in MRI. Yang XiaЧитать онлайн книгу.

Essential Concepts in MRI - Yang Xia


Скачать книгу
time-domain function is f0 in the frequency domain.

Function in time FT of the function in frequency
A sine or cosine function [e.g., sin(t)] A delta function at f
A constant [a DC offset with an amplitude] A spike at the origin
A square/rectangular pulse A sinc function [i.e., sin(θ)/θ]
A Lorentzian An exponential
A Gaussian A Gaussian

      2.8.2 Spectral Line Shapes – Lorentzian and Gaussian

      Spectral line shapes in NMR describe features of the energy exchange in an atomic system. As shown in Eq. (2.2), a nuclear transition is associated with a specific amount of energy, which would imply an extremely sharp spectral line in NMR. However, the spectral line as measured in NMR is not sharp but broadened considerably. The factors that broaden the spectral line include some fundamental physics principles as well as instrumentation factors.

      Figure 2.12 Comparison between a Lorentzian and a Gaussian with the same FWHM.

Lorentzian Gaussian
Line-shape equation f(x)=A(1+B2(x−x0)2 g(x)=e−(x−x0)2a2
Line width (FWHM) 2/B 2aIn(2)= 1.66511a
Normalized expression f1(x)=(B/πA)f(x) g1(x)=1xa2g(x)
Fourier transform Exponential Gaussian

      2.9 CW NMR

      The earliest NMR experiments ran in a continuous-wave (CW) mode, where the spectrometer is tuned to observe the component of M, which is 90˚ out of phase to the rotating field B1, the so-called absorption mode signal. (Earlier in Section 2.7, we set both B1(t) and u in the direction of x′, and v in the direction of y′ in the rotating frame.) During an experiment, the magnetic field B0 is swept slowly through the resonance frequency. As each chemically identical spin group comes into resonance, it undergoes nuclear induction and a voltage is induced in the pick-up coil (cf. the three peaks of ethanol in Figure 1.4). This approach is called the CW method, where the signal of the specimen is recorded continuously on an oscilloscope. Provided that this field sweep is done sufficiently slow, the absorption mode signal at each frequency corresponds to the steady state value of v when M has come to rest in the rotating coordinate system. Hence it is also called the slow passage experiment. Since neither the resonance frequency nor the number of the equivalent groups in a specimen is known, doing an NMR experiment using the CW method could take a long time.

      By examining the Bloch equation in the rotating frame [Eq. (2.23)], the following observations can be made:

      1 When we are far from the resonance (i.e., |ω0 – ω| is large), we have u = v = 0 and Mz = M0. The non-zero values of u and v appear only in a small interval around ω0, that is, when there is a resonance.

      2 Where T1T2(γB1)2 ≪ 1 (i.e., the rf power applied is sufficiently low so that the saturation does not occur), v can be simplified as (2.24)By comparing Eq. (2.24) with the line-shape functions in Table 2.3, we see that v is a Lorentzian centered at ω0 with a line width at half maximum of 1/(πT2). Hence, in principle, the FWHM of the resonant peak can be used to determine the T2 relaxation time.

      3 When T1T2(γB1)2 is not sufficiently smaller than 1, we can have these situations:when T1T2(γB1)2 < 1, the spins are below saturation, and the signal ∝


Скачать книгу
Яндекс.Метрика